IWATSUIWATSU Rogowski Coil Rogowski Datasheet datasheet.pdf
PD-200X Overview (V2.0).pdf
IWATSU rogowski IWATSU Rogowski Current Probe SS-280A Manual -Probemart.pdf
IWATSUIWATSU Rogowski Coil Rogowski Datasheet datasheet.pdf
IWATSU Rogowski Coil User Guide.pdf
IWATSU Current Probe Current Probe ?-Probemart.pdf
IWATSU Technical Material Rogowski Coil _ -Probemart.pdf
IWATSU Rogowski Coil Signal Droop Characteristics _ Technical Material _ -Probemart.pdf
IWATSU Application Article Rogowski Coil Overview -Probemart.pdf
IWATSU Application Article Rogowski Current Probe Principle -Probemart.pdf
Wide-Bandgap Device Dynamic Characterization iV1.03_2023_jectronic.pdf
The SS-280AH series uses an open flexible coil with a maximum wire diameter of 1.7mm, which can pass through pin testing. It is suitable for narrow and extremely small spaces, with a coil length of 84mm. Compared with traditional current sensors, it is easier to install and use, and can measure and monitor current online.
The SS-280A-H series has an appearance essentially the same as the SS-280A series, and the measurement range is also basically the same. The main difference is that the SS-280A-H series is resistant to high temperatures up to 150°C.
It is the only Rogowski coil that can withstand temperatures above 125°C.
————————————————————————————————————————————————————
Main features of the SS-280A-H series:
The probe tip diameter is as thin as 1.7mm and can directly detect TO-type MOSFETs and IGBTs used in switch-mode power supplies and other power electronics devices.
It has a frequency bandwidth of 30MHz, which allows observing high-frequency switching waveforms.
The current measurement range is 30A~30000A, supporting various current measurements. Since there is no magnetic saturation, large current measurements can be carried out with confidence.
The probe operating temperature range for A-H type is -40°C ~ 150°C
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
| Parameter | Value |
|---|---|
| Peak Current | 30000 A |
| High Frequency Bandwidth | 30 MHz |
| Low Frequency Cut-off | 2 Hz |
| Frequency Bandwidth | 2 Hz~30 MHz |
| Peak di/dt | 80 / 40 kA/µs (peak / continuous) |
| Sensitivity | 0.2 mV/A |
| Typical Accuracy | ±2% |
| Noise | 1.4 mV (rms, effective value) |
| Coil Withstand Voltage | 1.2 kV |
| Operating Temperature | -40~+150 °C |
| Zero Adjustment | Yes |
| Coil Diameter | Max 1.7 mm |
| Coil Length | 84±4 mm |