Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
IWATSU SS-500 series Hall-effect current IWATSU Catalog
IWATSU SS-500 series Hall-effect current probe AC/DC high-frequency clamp-on clamp min. measurable milliamp level
  • IWATSU SS-500 series Hall-effect current probe AC/DC high-frequency clamp-on clamp min. measurable milliamp level

IWATSU SS-500 series Hall-effect current probe AC/DC high-frequency clamp-on clamp min. measurable milliamp level

Brand: IWATSU

Origin: Japan

SS-500 Hall-effect AC/DC probe, current 0.5 A-700A, bandwidth 2-120MHz, measurement AC and DC clamp-on
Overview
Product Name IWATSU high-frequency Hall current clamp SS-500 series AC/DC current probe
Model SS- 531 530 521 520 570 560 550 540

Maximum

Rated current:

30A/5A/0.5A 5A 5A 500A 150A 30
Peak current: 50A/7.5A/0.75A 7.5A 7.5A 700A 300/500A 50
Bandwidth: 120MHz 50MHz 120MHz 50MHz 2MHz 10MHz 100MHz 50MHz


Specifications


参数项 SS-531 SS-530 SS-521 SS-520
Frequency bandwidth DC~120MHz (-3dB) DC~50MHz (-3dB) DC~120MHz (-3dB) DC~50MHz (-3dB)
Rise time (10~90%) 2.9ns or less 7.0ns or less 2.9ns or less 7.0ns or less
Max. rated current 30A range: 30Arms, 5A range: 5Arms, 0.5A range: 0.5Arms (DC, SINE)
54rms (DC, SINE)
Max. allowable peak current 30A range: ±50Apeak (2s or less, 10s for cool down is required.)5A range: ±7.5Apeak0.5A range: ±0.75Apeak (10mHz or less)±0.3Apeak (10mHz or higher)
±7.5 A peak (discontinuous)
Output rate 30A range: 0.1V/A5A range: 1V/A0.5A range: 10V/A
1 V/A
Amplitude accuracy 30A range: Typ. ±1.0%rdg. ±1mV, ±3.0%rdg. ±1 mV5A range: Typ. ±1.0%rdg. ±1mV, ±3.0%rdg. ±1 mV0.5A range: Typ. ±1.0%rdg. ±1mV, ±3.0%rdg. ±10 mV(DC, SINE 45~66 Hz, within the max. peak current value of each current range)
Typ. ±1.0%rdg. ±1mV, ±3.0%rdg. ±1 mV(DC, SINE 45~66 Hz, 0~54rms)
Noise level 75µArms or less (For only the probe with the 0.5 A range, connected with a measuring instrument that has a frequency band of 20 MHz)

Maximum rated power 7.8 VA (When measuring 30A ms continuously)
3.2 VA (With continuous maximum input)
Rated supply voltage DC ± 12V ± 0.5V

DC ± 12V ± 0.5V
Operating temperature and humidity range 0℃ ~ +40℃, 80%RH or less (no condensation) 0℃ ~ +40℃, 80%RH or less (no condensation)
Storage temperature and humidity range -10℃ ~ +50℃, 80%RH or less (no condensation) -10℃ ~ +50℃, 80%RH or less (no condensation)
Location for use Indoor, pollution degree 2, altitude up to 2000 m Indoor, pollution degree 2, altitude up to 2000 m
Effect of external magnetic fields 5mA or less(DC and 60 Hz, Magnetic field of 400 A/m) 20mA or less(DC and 60 Hz, Magnetic field of 400 A/m) 5mA or less(DC and 60 Hz, Magnetic field of 400 A/m) 20mA or less(DC and 60 Hz, Magnetic field of 400 A/m)
Diameter of measurable conductors φ 5mm or less φ 5mm or less
Cable lengths Sensor cable: 1.5m, Power cord: 1.0mBetween junction box and termination unit: 0.15m Sensor cable: 1.5m, Power cord: 1.0m
Dimensions Sensor: 155 (W) × 18 (H) × 26 (D) mmTermination unit: 29 (W) × 83 (H) × 40 (D) mmJunction box: 45 (W) × 120 (H) × 25 (D) mm Sensor: 155 (W) × 18 (H) × 26 (D) mmTermination unit: 29 (W) × 83 (H) × 40 (D) mm
Mass Approx. 370g Approx. 250g
Power supply (optional) PS-54 PS-54, PS-52
Standards Safety: EN61010, EMC: EN61326 Safety: EN61010, EMC: EN61326
Accessories Manual (1), Case (1)


参数项 SS-570 SS-560 SS-550 SS-540
Frequency bandwidth DC~2MHz (-3dB) DC~10MHz (-3dB) DC~10MHz (-3dB) DC~50MHz (-3dB)
Rise time (10~90%) 175ns or less 35ns or less 3.5ns or less 7.0ns or less
Max. rated current 500A 150A 30Arms
Max. allowable peak current 700Apeak (discontinuous) 300Apeak (discontinuous)
500Apeak (Pulse width ≤30µs)
50Apeak (discontinuous)
Output rate 0.01 V/A 0.1 V/A
Amplitude accuracy ~500A ±1.0%rdg. ±5mV,
~700Apeak ±2.0%rdg.
(DC, SINE 45~66 Hz, 25Arms)
~150A ±1.0%rdg. ±5mV,
~300Apeak ±2.0%rdg.
(DC, SINE45~66 Hz, 25Arms)
~30A ±1.0%rdg. ±1mV,
~50Apeak ±2.0%rdg.
(DC, SINE45~66 Hz)
Noise level 2.5mArms or less (for measuring instrument with 20 MHz bandwidth) 2.5mArms or less (for measuring instrument with 20 MHz bandwidth)
Maximum rated power 7.2VA (with continuous maximum input) 5.5VA (with continuous maximum input) 5.3VA (with continuous maximum input) 5.6VA (with continuous maximum input)
Rated supply voltage DC ±12V±0.5V DC ±12V±1V
Operating temperature and humidity range 0℃ ~ +40℃, 80%RH or less (no condensation) 0℃ ~ +40℃, 80%RH or less (no condensation)
Storage temperature and humidity range -10℃ ~ +50℃, 80%RH or less (no condensation) -10℃ ~ +50℃, 80%RH or less (no condensation)
Location for use Indoor, pollution degree 2, altitude up to 2000 m Indoor, pollution degree 2, altitude up to 2000 m
Effect of external magnetic fields 800mA or less
(DC and 60 Hz, Magnetic field of 400 A/m)
150mA or less
(DC and 60 Hz, Magnetic field of 400 A/m)
5mA or less
(DC and 60 Hz, Magnetic field of 400 A/m)
20mA or less
(DC and 60 Hz, Magnetic field of 400 A/m)
Diameter of measurable conductors φ 20 mm or less φ 5 mm or less
Cable lengths Sensor cable: 1.5m, Power cord: 1.0m Sensor cable: 2.0m, Power cord: 1.0m
Dimensions Sensor: 176 (W) × 69 (H) × 27 (D) mm
Termination unit: 27 (W) × 55 (H) × 18 (D) mm
Sensor: 175 (W) × 18 (H) × 40 (D) mm
Termination unit: 27 (W) × 55 (H) × 18 (D) mm
Mass Approx. 520g Approx. 500g Approx. 240g Approx. 230g
Power supply (optional) PS-54, PS-52
Standards Safety: EN61010, EMC: EN61326 Safety: EN61010, EMC: EN61326
Accessories Manual (1), Case (1)



Expand IWATSU SS-500 series Hall-effect current

* Specifications are subject to change without notice.