Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Magnelab high-frequency pulse current tr MagneLab Catalog
Magnelab high-frequency pulse current transformer CT-D0.25 ultra-fast RF FCT high-current probe EMC test medium-aperture 4000 A/100M
  • Magnelab high-frequency pulse current transformer CT-D0.25 ultra-fast RF FCT high-current probe EMC test medium-aperture 4000 A/100M

Magnelab high-frequency pulse current transformer CT-D0.25 ultra-fast RF FCT high-current probe EMC test medium-aperture 4000 A/100M

Brand: MagneLab

Origin: USA

CT-D0.25, aperture 1, peak current 4000 A, high-frequency bandwidth 100M
Overview

Magnelab ultra-high-speed current transformers (FCT) enable precise isolated measurement of unipolar and bipolar pulses or AC currents.

This product series measures currents from microampere level to 20 kA over a frequency range of 0.5 Hz to 500 MHz.
The FCT output signal is a voltage waveform at a defined transformation ratio of the measured current and can be analyzed on an oscilloscope, RF power meter, spectrum analyzer, or custom interface circuit.
Magnelab FCTs have excellent performance, are offered in 7 ranges and 5 housing sizes, and can be terminated with SMA or BNC connectors.
All FCTs can be supplied with a calibration certificate and factory test data to ensure optimum performance.
For special application requirements, Magnelab can provide customized FCT ultra-high-speed current transformers to meet user requirements.
The FCT can measure pulses with rise times down to below 200 ps (picoseconds) and bandwidths exceeding 2 GHz, as well as continuous pulse-burst signals.
As an ultra-high-speed current transformer, the Magnelab FCT is the current sensor with the highest sensitivity and fastest response worldwide.

______________________________________________________________________________

Main advantages of Magnelab current transformers:20200303095348958.jpg

• Higher frequency bandwidth ensures true representation of fast transients
• 5× higher sensitivity for small-signal measurement
• Operating temperature -20°C to +120°C
• Accuracy of ±0.5% across the full frequency band

______________________________________________________________________________

Application areas and industries include:

• Laser and plasma research
• EMC and EMI research
• Semiconductor gate switching
• Power system harmonics and transients
• Lightning research and simulated testing
• Capacitors and electrostatic discharge
• Corona wire discharge
• Partial discharge measurement
• Strictly monitor current fluctuations


Specifications

115ef93a-f353-40f1-8572-62074d0d07c2.png

Model Output (V/A) in 1 MΩ RMS(A) Maximum peak (A) Droop (%/us) Rise time (ns) Max l.t*(As)in 50Ω -3db low(Hz) -3db high(MHz)
CT-B5.0 5 2 200 3 0.875 0.0001 4800 400
CT-B2.5 2.5 5 400 0.75 0.7 0.0004 1200 500
CT-B1.0 1.0 8 1000 0.13 0.7 0.0025 200 500
CT-B0.5 0.5 11 2000 0.03 1.75 0.01 48 200
CT-B0.25 0.25 16 4000 0.015 3.5 0.013 24 100
CT-B0.1 0.1 25 10000 0.006 7 0.013 10 50
CT-B0.05 0.05 35 20000 0.003 17.5 0.053 5 20
CT-C5.0 5 2 200 3 0.875 0.0002 4800 400
CT-C2.5 2.5 5 400 0.75 0.7 0.0008 1200 500
CT-C1.0 1.0 11 1000 0.13 0.7 0.005 200 500
CT-C0.5 0.5 16 2000 0.03 1.75 0.02 48 200
CT-C0.25 0.25 22 4000 0.01 3.5 0.08 12 100
CT-C0.1 0.1 35 10000 0.004 7 0.167 6 50
CT-C0.05 0.05 50 20000 0.002 17.5 0.283 3 20
CT-D5.0 5 2 200 3 0.875 0.0002 4800 400
CT-D2.5 2.5 5 400 0.75 0.7 0.0008 1200 500
CT-D1.0 1.0 11 1000 0.13 0.7 0.005 200 500
CT-D0.5 0.5 16 2000 0.03 1.75 0.02 48 200
CT-D0.25 0.25 22 4000 0.01 3.5 0.08 12 100
CT-D0.1 0.1 35 10000 0.002 7 0.5 2 50
CT-D0.05 0.05 50 20000 0.001 17.5 0.756 1 20
CT-E5.0 5.0 2 200 3 0.875 0.0002 4800 400
CT-E2.5 2.5 5 400 0.75 0.7 0.0008 1200 500
CT-E1.0 1.0 11 1000 0.13 0.7 0.005 200 500
CT-E0.5 0.5 22 2000 0.03 1.75 0.02 48 200
CT-E0.25 0.25 32 4000 0.01 3.5 0.08 12 100
CT-E0.1 0.1 50 10000 0.002 7 0.5 2 50
CT-E0.05 0.05 71 20000 0.001 17.5 0.756 1 20









CT-F5.0 5.0 2 200 3 0.875 0.0002 4800 400
CT-F2.5 2.5 5 400 0.75 0.7 0.0008 1200 500
CT-F1.0 1.0 11 1000 0.13 0.7 0.005 200 500
CT-F0.5 0.5 22 2000 0.03 1.75 0.02 48 200
CT-F0.25 0.25 32 4000 0.01 3.5 0.08 12 100
CT-F0.1 0.1 50 10000 0.002 7 0.5 2 50
CT-F0.05 0.05 71 20000 0.001 17.5 0.756 1 20
CT-CALCERT Certificate of calibration with amplitude vs.frequency response plots on option


Expand Magnelab high-frequency pulse current tr

* Specifications are subject to change without notice.