The TR9310A can measure DC voltages up to 10KV and AC voltages up to 7KV, with an input impedance of 100MΩ and low loading effect. Its high-frequency bandwidth of 120MHz allows you to capture fast, high-voltage signals. The high-voltage probe can be connected directly to an oscilloscope for measurement, making it suitable for high-voltage measurements referenced to ground. Additionally, with an optional adapter, the voltage value can be measured on a digital multimeter. The TR9310A uses an environmentally friendly silicone resin insulating medium, which enhances insulation performance, ensures that internal components do not oxidize during high-voltage measurements, and extends the service life.
Product Features
• Maximum measurable voltage: 10KV DC, ACpeak
• Bandwidth up to 120MHz
• Input impedance 100MΩ
• Fixed attenuation ratio 100:1, Read Out
• Can be connected to a multimeter using an impedance matcher
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
| Model | TR9310A |
| attenuation ratio | 100:1 |
| Read aloud function) | O |
| Input impedance | 100 MΩ |
| Input capacitance | 2.0PF |
| Maximum Input: DC AC peak | 10KV CAT II |
| Maximum Input: AC RMS | 7KV CAT II |
| Pulse Voltage | |
| Maximum load current | 100 µA |
| Bandwidth | DC~120MHz |
| rise time | 3.0 nS |
| Noise ratio | >60dB at 1KHz; >50dB at 1MHz |
| DC signal error value | ≤3% full range scale, ≤3% (0-35KV) |
| AC signal error value | ≤3% at 1KHz |
| Temperature coefficient | ≤200 PPM/℃ |
| Matching capacitor adjustment range | 10PF~35PF |
| Safety Standard Inspection | Complies with EN61010-031 CAT II standard |
| Test line length | 2M ± 0.2M |
| Operating Temperature and Humidity | 0~40℃, 80% or less |
| Save temperature and humidity | -10~45℃, 80% or less |
| Color: Grip/Body | Black/Light Gray |
| Size | 80mm x 80mm x 320mm |
| Weight | 460g |