Rogowski Coil Low-Frequency and Droop Characteristics
Source: Probemart 2024-12-11 Views: 1665
1. Overview
Description
In principle, a Rogowski coil current probe cannot perform DC testing. Therefore, it has the characteristic that detection sensitivity decreases at low frequencies. Generally speaking,
The low-frequency cutoff frequency is defined as the frequency at which detection sensitivity drops to -3 dB.
Affected by this low-frequency characteristic, when testing signals containing low-frequency components, the actual current and the result measured by the Rogowski probe will differ to some extent; pay close attention during use.
Low-frequency cutoff frequency and roll-off specifications for each model
Another low-frequency characteristic specification related to the low-frequency cutoff frequency is roll-off (Droopping). The roll-off referred to here is the characteristic when testing a step pulse waveform,
The basic specification values are shown in Table 1 below.
Table 1 Basic specifications of roll-off
Model |
Low-frequency cutoff frequency [Hz] |
Roll-off [%/ms] |
SS-281A |
110 |
80 |
SS-282A |
65 |
50 |
SS-283A |
32 |
25 |
SS-284A |
9 |
9 |
SS-285A |
6 |
6 |
SS-286A |
3 |
3 |
SS-287A |
2 |
2 |
SS-293S/L |
1 |
0.8 |
SS-294S/L |
0.8 |
0.7 |
SS-295S/L |
0.6 |
0.5 |
SS-296S/L |
0.4 |
0.35 |
Definition and characteristics of roll-off
Roll-off (Droopping) is the amount of attenuation after a certain time when the pulse waveform rises to the top.
When a Rogowski coil probe is used, it is defined as the attenuation after 1 ms; the unit is [%/ms].
This ratio increases as the low-frequency cutoff frequency increases.

2. Effect on the measured waveform
The following is a comparison of test results between the DC current probe (SS-240A) and a Rogowski current probe.
It can be seen that the lower the low-frequency cutoff frequency, the smaller the effect of roll-off; at the same time, the lower the sensitivity (the larger the detection range), the more noise is caused
the greater the effect. Therefore, be sure to select the appropriate model according to the actual test purpose.



2.4. Switching waveform (pulse adjustable)
Items ①–③ below are switching waveforms simulated using a variable-frequency circuit (fixed pulse width); the switching frequency is set to 1.8kHz and the AC signal frequency to 50Hz.
The switching frequency is already sufficiently high relative to the low-frequency cutoff frequency, but because an AC signal component (50Hz) is included, its effect on the measured waveform can be seen.
① UsingCase of SS-283A (low-frequency cutoff frequency 32Hz)




PCIM Europe 2026, Nuremberg: IWATSU and PMK at Booth H7-157
2026-06-15
APEC 2026: 41st IEEE Applied Power Electronics Conference and Exposition
2026-03-15
Semicon China 2026, Shanghai New International Expo Centre (SNIEC)
2026-03-11
WiPDA Asia 2025, Beijing International Convention Center, 15 August
2025-08-06